CRA · CASE STUDY

Eighty units still await remediation

Of 800 affected devices, 720 are confirmed.

PRODUCT
CONFORMITY
MAINTENANCE

Starting position and test order

A manufacturer detects a product-related protection deviation in a batch of 800 devices in the fictitious case. 200 units are located at distribution partners and 600 are already with end users. For the application case, an effective update is possible. For other units the team assesses separately necessary measures.

180 sales units are proven to be withheld. 540 end-user devices clearly demonstrate the application of the verified correction. For 20 distribution units and 60 end-user devices, the implementation documentation is missing. All eighty units can be identified using their delivery data.

Compare test data and consequences

  • Distribution

    Scale
    200 specific units
    Observation
    180 restraint measures confirmed
    Follow
    20 measures to demonstrate
  • Users

    Scale
    600 units concerned
    Observation
    540 corrections confirmed
    Follow
    60 follow-up actions required
  • Total population

    Scale
    800 clearly identified units
    Observation
    720 effective measures demonstrated
    Follow
    80 units open

Confirmed return and open stock

Out of 200 distribution devices, 180 are reached. 540 corrections are confirmed for 600 end-user devices. A total of 80 are open.
Out of 200 distribution devices, 180 are reached. 540 corrections are confirmed for 600 end-user devices. A total of 80 are open.

Count unique devices per inventory and continue to process the open 20 sales and 60 end-user devices.

Edit the test step by step

  1. Set the population and measures per stock

    Arrange sales inventory and installed devices using unique identifiers. Set effective measures per section.

    Testable evidence A traceable list of 200 and 600 devices with a reasonable approach.

    Acceptance criterion
    Reconcile stock in distribution and installed units separately to the affected release. Check serial numbers for duplicates.
    Evidence
    Measurement record with baseline population, source, expected and observed values, unit and timestamp.
  2. Matching the confirmed effect and return

    Count a verifiable performance per unit. Check identity, effectiveness and delimitation of notifications separately.

    Testable evidence 720 documented measures and 80 open units.

    Acceptance criterion
    Count each unit with a confirmed action once. Retain evidence of execution and observed effect for that unit.
    Evidence
    Copy of the submitted version with version reference, dispatch evidence, acknowledgement and outstanding items.
  3. Managing the residual portfolio risk-appropriate

    For 20 distribution devices, record the security to be demonstrated and for 60 user devices a suitable, scheduled follow-up action. Evaluate withdrawal or recall if necessary.

    Testable evidence Released or kept open measure with residual list and escalation.

    Acceptance criterion
    For each open unit, record the agreed action, who must evidence it and when the remaining population will be reviewed.
    Evidence
    Risk entry with impact, action, residual risk and approval.

Worked decision and negative test

Decision in case

The remedy is documented in the sample of 720 out of 800 devices. The open quota does not replace risk assessment. The next steps are based on the location and effectiveness of achievable measures.

Keep the case open until the remaining units are adequately treated and the required supporting documents are secured. Vote for any official measures with the competent authority.

Counter-sample and impact

Complete twenty telephone contact attempts without delivery or implementation documents. Communication is better documented. The number of devices with proven success does not change. The open inventory continues to include 80 devices.

Audit record
Comparison of the original finding and retest showing the changed condition, observed effect and closure decision.

Templates and worksheets

  • Tracking measures and open equipment stock

    Arrange effectively confirmed measures to clear units and their distribution or usage status.

    Outcome Residual list of responsibility, action, proven impact and next target date.

    Read the instructions

    Fictional example 720 measures are taken by 800 affected units. 20 distribution devices and 60 user devices remain open.

    1. Identify the starting situation Determine clearly matched stock according to actual location.
    2. Run the test Effectively confirmed measures distinguish from contact attempts.
    3. Secure decision and evidence Edit the remaining stock and prove the final decision.

Official sources

What depends on your result?

  • Track correction, withdrawal and recall

    Transfer the same test criteria to your own product and save the different starting data.

    View the guide
  • Processing market surveillance request

    Prepare the feedback to an authority with a confirmed scope, action and open balance.

    View the guide
  • Release laboratory

    Check how a double entry changes the decision on the remaining stock.

    Start exercise